Philips Ultrasound has issued a letter to affected customers recommending all Lumify diagnostic ultrasound systems have updated use instructions, according to an FDA Early Alert.   

The FDA reports that due to a software issue, the Lumify Android or iOS application used with its ultrasound transducers may not recognize a previously registered transducer and will prompt users for re-registration without warning prior to point-of-care use. Scanning functionality is prevented until the registration step has been completed. Registration requires Wi-Fi or cellular signal access. 

Lumify devices used outside the hospital, such as in ambulances, aircraft, or field settings, are at greater risk due to more limited access to Wi-Fi/cellular signals. Philips has received reports that the unexpected re-registration events have prevented scanning in pre-hospital environments and delayed care to patients. These pre-hospital events occurred outside of Wi-Fi or cellular access. 

Additionally, Philips reported that the connection between the Lumify transducer USB cable to a customer’s tablet or phone may disconnect or have an unstable connection during use, which may disrupt continuous function of the device and result in a delay of care. 

According to the FDA, loss of functionality or system failure during a procedure may result in a delay in diagnosis and/or treatment. Because the Lumify system is intended for point-of-care use, where ultrasound imaging supports rapid diagnosis and treatment; real-time decision-making; triage of unstable patients; and access to diagnostic capabilities in mobile, underserved, or remote settings, loss of functionality or system failure may have greater consequences than with traditional ultrasound devices, the FDA reports.

Depending on the clinical situation and patient condition, delays in diagnosis or treatment could lead to serious injury or death, according to the FDA.

As of September 29, 2026, Philips Ultrasound has reported eight serious injuries and four deaths associated with this issue, the FDA reports.

Affected Devices

Software Version    UDI-DI (GTIN)    
Lumify Diagnostic Ultrasound System iOS (2.X)00884838103405
Lumify Diagnostic Ultrasound System iOS (5.0)00884838120266
Lumify Diagnostic Ultrasound System iOS (5.1.2)00884838124141
Lumify Diagnostic Ultrasound System iOS (5.1.3)00884838124141
Lumify Diagnostic Ultrasound System Android (1.X)00884838106987
Lumify Diagnostic Ultrasound System Android (3.X)00884838100121
Lumify Diagnostic Ultrasound System Android (4.X)00884838108455
Lumify Diagnostic Ultrasound System Android (5.1)00884838122116
Lumify Diagnostic Ultrasound System Android (5.1.1)00884838122116


FDA Recommendations

To prevent unintended loss or delays in scanning functionality, only use the Lumify diagnostic ultrasound system in areas with readily accessible Wi-Fi/cellular signals unless a backup ultrasound device is immediately available.

On September 24, 2026, Philips Ultrasound sent all affected customers a letter recommending the following actions:

  • To prevent unintended loss or delays in scanning functionality, only use the Lumify diagnostic ultrasound system in areas with readily accessible Wi-Fi/cellular signals unless a backup ultrasound device is immediately available.
  • Always have backup ultrasound devices and/or modalities available for care.
  • Circulate the Philips Ultrasound provided letter to all users of the device so that they are aware of the product issue and its associated hazard/harm.
  • Post the notice provided by Philips Ultrasound near the affected unit(s) for ease of reference

More information is available at the FDA website.