ICU Medical has issued a letter to affected customers recommending certain endotracheal tubes be removed from where they are used or sold, according to an FDA early alert.

Portex Endotracheal Tubes are intended for oral and/or nasal intubation for airway management. Portex Polar preformed endotracheal tubes are intended for airway management during surgical procedures involving the head, neck or mouth where it would be advantageous to remove all connections from the operative field, the FDA says.

According to the FDA, ICU Medical has identified a manufacturing issue affecting certain lots of Portex Endotracheal Tubes. Affected endotracheal tubes were manufactured with a resin that has variability in the percentage of plasticizer, which may cause endotracheal tube material to be softer than expected.

Use of affected endotracheal tubes intended for the neonatal and early infant population may lead to a delay in intubation due to increased resistance to ventilation, tube kinking, and difficulty to intubate due to material softness. If the issue is not identified during intubation, the issue may result in inadequate ventilation and unscheduled tube replacement procedures that may lead and/or contribute to serious adverse events including death.

As of November 16, 2026, ICU Medical has reported one death associated with this issue, according to the FDA alert.

FDA Recommendations

Check all inventory for affected product and immediately discontinue use. Destroy products or segregate and quarantine if immediate destruction is not possible.

On November 20, 2026, ICU Medical sent all affected customers a letter recommending the following actions:

  • Check all inventory locations within your institution for the affected products and immediately discontinue use. 
  • Destroy all affected products following your institution’s process for destruction.
    • If destruction of affected products is not immediately possible at your facility, segregate and quarantine all affected products to prevent further distribution or use. 
  • Post ICU Medical’s communication on or near the affected product until destruction is complete.
  • Immediately forward ICU Medical’s notification and the attached Response Form to all customers and users to whom you distributed these products to ensure they are aware of this notification.
  • Report any adverse health complaints experienced with the use of this product to ICU Medical.

More information is available on the FDA website.