Medline Industries has recalled specific lots of its bubble CPAP kits because of potential heating failures in the breathing circuit.
RT’s Three Key Takeaways:
- Thermal Damage Risk: Medline Industries recalled infant bubble continuous positive airway pressure kits after reports that heated wire connectors could melt, smoke, or show signs of localized heat damage.
- Affected Inventory: The Class 2 recall affects 24 units under SKU HUD26007KIT and lot number 25DBT748, which were distributed to facilities in Florida and Texas.
- Required Facility Action: Healthcare consignees must destroy affected inventory and submit a verification form to Medline to receive credit.
Medline Industries has recalled specific lots of its bubble continuous positive airway pressure (bCPAP) kits because of potential heating failures in the breathing circuit, according to an FDA alert.
The FDA classified the event as a Class 2 device recall on Nov 20, 2026, after Medline first initiated the correction on Oct 7, 2026. Class 2 recalls are designated when exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
According to the FDA notice, neonatal and infant heated wire breathing circuits included in the kits may experience sporadic operational failure. Medline reported that the heated wire connector pigtail at the humidifier can exhibit discoloration, melting, smoke, a burning odor, or other signs of localized thermal damage.
The federal agency noted that the root cause remains under investigation by the firm.
Affected Products
The recall covers 24 units distributed within the US, specifically to consignees in Florida and Texas, according to regulatory data.
Healthcare providers and respiratory therapy departments should inspect stock for the following product identifiers:
- Product Name: Medline BUBBLE CPAP KIT
- Medline Kit SKU: HUD26007KIT
- Lot Number: 25DBT748
- UDI/DI (Each and Case): 40195327227198
- Recall Number: Z-0523-2027
FDA Recommendations
Medline issued an urgent medical device correction notice to affected customers via email and postal mail in October 2026, according to the recall database. The communication explained the thermal risk, outlined patient safety concerns, and requested that all affected kits be destroyed immediately.
The manufacturer stated that credit will be issued to facilities once a completed response form is received. Consignees with questions can reach Medline’s recall department by telephone at 866-359-1704 or via email at [email protected].